A Palladium-Tin Modified Microband Electrode Array for Nitrate Determination

机译:用于硝酸盐测定的钯锡修饰微带电极阵列

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A microband electrode array modified with palladium-tin bimetallic composite has been developed for nitrate determination. The microband electrode array was fabricated by Micro Electro-Mechanical System (MEMS) technique. Palladium and tin were electrodeposited successively on the electrode, forming a double-layer structure. The effect of the Pd-Sn composite was investigated and its enhancement of catalytic activity and lifetime was revealed. The Pd-Sn modified electrode showed good linearity (R2 = 0.998) from 1 mg/L to 20 mg/L for nitrate determination with a sensitivity of 398 μA/(mg∙L−1∙cm2). The electrode exhibited a satisfying analytical performance after 60 days of storage, indicating a long lifetime. Good repeatability was also displayed by the Pd-Sn modified electrodes. The results provided an option for nitrate determination in water.
机译:已经开发了用钯-锡双金属复合材料修饰的微带电极阵列,用于硝酸盐的测定。该微带电极阵列是通过微机电系统(MEMS)技术制造的。钯和锡连续地电沉积在电极上,形成双层结构。研究了Pd-Sn复合材料的作用,并揭示了其催化活性和寿命的提高。 Pd-Sn修饰电极在1 mg / L至20 mg / L的硝酸盐测定中显示出良好的线性(R 2 = 0.998),灵敏度为398μA/(mg∙L - 1 ∙cm 2 )。储存60天后,电极表现出令人满意的分析性能,表明使用寿命长。 Pd-Sn修饰电极也显示出良好的重复性。结果为水中硝酸盐的测定提供了一种选择。