Comparison of corneal thickness measurements using Orbscan II non-contact specular microscopy and ultrasonic pachymetry in eyes after laser in situ keratomileusis

机译:激光原位角膜磨镶术后使用Orbscan II非接触式镜面显微镜和超声测厚法测量角膜厚度的比较

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

>Aims: To compare central corneal thickness measurements of three pachymetry devices in eyes after laser in situ keratomileusis (LASIK).>Methods: Central corneal thickness was measured in 203 eyes after myopic LASIK. Orbscan II scanning slit topography (Bausch & Lomb), SP-2000P non-contact specular microscopy (Topcon), and ultrasonic pachymetry (Tomey) were used in this sequence.>Results: Three devices gave significantly different corneal thickness readings (p<0.0001, repeated measure analysis of variance). The measurements of Orbscan II (445.6 (SD 60.0) μm) were significantly smaller than those of noncontact specular microscopy (467.9 (SD 40.2) μm; p<0.0001, Tukey multiple comparison) and ultrasonic pachymetry (478.8 (SD 41.9) μm; p<0.0001). The value obtained with SP-2000P non-contact specular microscopy was significantly smaller than that taken with ultrasonic pachymetry (p<0.001). There were significant linear correlations between scanning slit topography and non-contact specular microscopy (Pearson’s correlation coefficient r = 0.912, p<0.0001), non-contact specular microscopy and ultrasonic pachymetry (r = 0.968, p<0.0001), and ultrasonic pachymetry and scanning slit topography (r = 0.933, p<0.0001).>Conclusion: In post-LASIK eyes, Orbscan II scanning slit topography significantly underestimated corneal thickness. Non-contact specular microscopy gave smaller thickness readings than ultrasonic pachymetry, but these two units showed an excellent linear correlation.
机译:>目的:比较激光原位角膜磨镶术(LASIK)后三种测厚仪在眼睛中的中央角膜厚度测量结果。>方法:在近视LASIK术后203眼中测量中央角膜厚度。依次使用Orbscan II扫描狭缝形貌(Bausch&Lomb),SP-2000P非接触式镜面显微镜(Topcon)和超声测厚仪(Tomey)。>结果:三种设备产生的角膜明显不同厚度读数(p <0.0001,重复测量方差分析)。 Orbscan II(445.6(SD 60.0)μm)的测量值显着小于非接触式镜检显微镜(467.9(SD 40.2)μm; p <0.0001,Tukey多次比较)和超声测厚法(478.8(SD 41.9)μm; p <0.0001)。 SP-2000P非接触式镜面显微镜获得的值明显小于超声测厚法(p <0.001)。扫描狭缝形貌与非接触式镜检显微镜(Pearson相关系数r = 0.912,p <0.0001),非接触式镜检显微镜和超声测厚法(r = 0.968,p <0.0001),超声测厚法之间存在显着的线性相关性。扫描狭缝形貌(r = 0.933,p <0.0001)。>结论:在LASIK术后眼睛中,Orbscan II扫描狭缝形貌显着低估了角膜厚度。非接触式镜面显微镜比超声波测厚仪具有更小的厚度读数,但是这两个单元显示出极好的线性相关性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号