Patellar mobility can be reproducibly measured using ultrasound

机译:ultrasound骨活动度可以使用超声波重现

获取原文
           

摘要

The present study was performed to examine the reliability of ultrasound in evaluating patellar mobility in the superior–inferior direction. Twelve healthy men volunteered for the study. Patellar mobility in the superior–inferior direction during isometric knee extension contraction with the knee immobilized in a semi-flexed knee brace was measured using ultrasound. Both intra-observer and inter-observer reliability were assessed by intra-class correlation coefficients (ICCs). Bland–Altman analysis was used for assessing agreement between measurements. ICC values were excellent for both intra-observer and inter-observer reliability at 0.97 and 0.93, respectively. In 95 % of measurements, the same observer measured within −0.55 to 0.61 mm, while different observers measured within −0.82 to 0.85 mm. In conclusion, patellar mobility in the superior–inferior direction during an isometric knee extension exercise can be reproducibly measured using ultrasound. VI (basic study of a novel evaluation method).
机译:本研究旨在检查超声在评估上下方向pa骨活动度方面的可靠性。十二名健康男人自愿参加了这项研究。使用超声波测量在等距膝关节伸展收缩过程中膝盖固定在半屈曲的膝盖支架中时the骨在上-下方向的活动性。观察者内部和观察者之间的可靠性均通过类内相关系数(ICC)进行评估。使用Bland–Altman分析来评估测量之间的一致性。 ICC值对于观察者内部和观察者之间的可靠性均极佳,分别为0.97和0.93。在95%的测量中,同一观察者的测量结果在-0.55至0.61毫米范围内,而不同的观察者的测量结果在-0.82至0.85毫米范围内。总之,在等距膝盖伸展运动中,pa骨在上下方向上的活动性可以使用超声波重现。 VI(一种新型评估方法的基础研究)。