A noninvasive tissue current measurement technique and its use in measuring a nonuniform current density are described. This current density image is created by measuring the magnetic field arising from these currents and taking its curl. These magnetic fields are proportional to the phase component of a complex magnetic resonance image. Measurements of all three components of a quasistatic nonuniform current density in a phantom are described. Expected current density calculations from a numerical solution for the magnetic field which was created by the phantom are presented for comparison. The results of a numerical simulation of the experiment, which used this field solution and which included the effects of slice selection and sampling, are also presented. The experimental and simulated results are quantitatively compared. It is concluded that the principle source of systematic error was the finite slice thickness, which causes blurring of boundaries.
机译:描述了一种非侵入性组织电流测量技术及其在测量非均匀电流密度中的用途。该电流密度图像是通过测量由这些电流产生的磁场并使其弯曲来创建的。这些磁场与复杂的磁共振图像的相位分量成比例。描述了幻影中准静态不均匀电流密度的所有三个分量的测量。提出了由幻影创建的磁场数值解的预期电流密度计算,以进行比较。还给出了使用该现场解决方案并包括切片选择和采样效果的实验数值模拟结果。实验结果和模拟结果进行了定量比较。结论是系统误差的主要来源是有限的切片厚度,这会导致边界模糊。