Like many oxidants, ozone is a volatile substance that can be difficult to measure accurately. One major cause of analytical error is ozone loss during sampling. The practices described below can improve ozone results. When designing the sampling system, engineers must keep in mind that ozone is aggressive and reactive and has a relatively short half-life, so construction materials are important. The sampling system should include 1. a tap that ensures samples are free of bubbles, sediment, and other impurities: 2. sample lines made of appropriate corrosion-resistant materials that are as short and as small in diameter as practical to minimize volume and detention time; 3. a mechanism that keeps sample lines free of accumulated solids; 4. sufficient inflow (sample volume) to the analyzer; and 5. a means of collecting grab samples near the on-line analyzer without disturbing analyzer inflow.
机译:像许多氧化剂一样,臭氧是一种挥发性物质,可能难以准确测量。分析误差的主要原因之一是采样过程中的臭氧损失。下述做法可以改善臭氧效果。在设计采样系统时,工程师必须记住,臭氧具有腐蚀性和反应性,并且半衰期相对较短,因此建筑材料非常重要。采样系统应包括1.确保样品无气泡,沉淀物和其他杂质的水龙头:2.采样管由合适的耐腐蚀材料制成,其长度尽可能短,直径要尽可能小时间; 3.一种使样品管线中没有累积的固体的机制; 4.足够的流入量(样品量)到分析仪; 5.一种在在线分析仪附近收集抓取样品而不会干扰分析仪流入的方法。