An unconstrained surface metacarpophalangeal joint replacement was developed with metal metacarpal and polyethylene phalangeal components, fixed by uncemented finned polyethylene plugs which allowed some metacarpal component motion. Clinical and radiological results in 13 joints in eight patients are presented after 5 years. One infection required revision at 3 months. There were no further complications. At final review there were no implant failures. Two of 13 joints showed lucency around the phalangeal component and one showed 2mm subsidence of the metacarpal component. No other adverse radiological features were observed. Seven patients had no pain and one had minor discomfort. Joint movement had improved from an arc of 27 degrees to 60 degrees and disability, assessed using the P.E.M. questionnaire, had improved from 77% to 9%.
机译:用金属掌骨和聚乙烯指骨组件开发了不受约束的表面掌指关节置换术,并通过未粘骨的有鳍聚乙烯栓固定,从而允许一些掌骨组件运动。五年后,在八名患者的13个关节中获得了临床和放射学结果。一种感染需要在3个月时进行修复。没有其他并发症。在最终审查中,没有植入失败。 13个关节中有2个在指骨组件周围显示出通透性,其中1个显示了掌骨组件下沉2mm。没有观察到其他不利的放射学特征。 7例患者无疼痛,其中1例轻微不适。使用P.E.M.评估,关节运动从27度的弧度提高到60度,并改善了残障能力。问卷,从77%提高到9%。