摘要

We would like to thank Mr. Perkins for his comments on the "eFlow study."He adds references to discuss aspects of this article, which partly give additional data, but do not contradict ours, as our eFlow study is designed to investigate changes in performance after 6 months of daily clinical practice. From that study it is obvious that people use different drugs in the same device (based on questionnaire data), although instructed not to do so. Nor do they clean their devices according to manufacturer's instructions. As explained in our article, we had to extend the study with a second series of used devices after it became clear that the devices in the first series were not maintained properly. This conclusion was based on checking the nebulizers upon pollution with drug residues.
机译:我们要感谢Perkins先生对“ eFlow研究”的评论。他添加了一些参考资料来讨论本文的各个方面,部分提供了其他数据,但并不与我们的数据相抵触,因为我们的eFlow研究旨在调查eFlow研究的变化。每天临床实践6个月后的表现。从该研究中可以明显看出,尽管有人指示不要在同一设备上使用不同的药物(基于调查表数据)。他们也没有按照制造商的说明清洁设备。正如我们在文章中所解释的,我们必须在发现第一系列中的设备维护不当之后,才将研究范围扩大到第二系列中的二手设备。该结论基于检查喷雾器是否受到药物残留污染。