Analytical results of anion determination by suppressed ion chromatography are significantly affected by calibration curve calculation. In this paper, as expected, eluent pK(a) is shown to influence calibration linearity in the range 1-20 mg/l sulfate, with A carbonate-hydrogencarbonate mixture producing a larger non-linearity than NaOH. Evidence is given for very large errors (about 30-40%) in estimating sample sulfate concentration when linear regression is used instead of a quadratic calibration curve. This study was performed following a 2(4) run full factorial experimental design, including eluent pK(a), counterion type, solution composition and current level for background suppression as main variables.
机译:通过抑制离子色谱法测定阴离子的分析结果受校准曲线计算的影响很大。在本文中,正如预期的那样,显示出洗脱液pK(a)在1-20 mg / l硫酸盐范围内影响校准线性,碳酸氢盐与碳酸氢盐的混合物产生的非线性比NaOH大。当使用线性回归而不是二次校准曲线估算硫酸盐浓度时,存在很大的误差(约30-40%)。该研究是按照2(4)运行的全因子实验设计进行的,其中包括洗脱液pK(a),抗衡离子类型,溶液成分和用于背景抑制的电流水平作为主要变量。