Pachymetry assessment with Orbscan II in postoperative patients with myopic LASIK.

机译:Orbscan II对近视LASIK术后患者进行测厚法评估。

获取原文
获取原文并翻译 | 示例
       

摘要

PURPOSE: To compare central corneal thickness after LASIK for myopia, using ultrasonic pachymetry and Orbscan II measurements, and to evaluate changes in these measurements over time. METHODS: Central corneal thickness measurements obtained by ultrasonic pachymetry and Orbscan II (Bausch and Lomb, Rochester, NY) in patients who underwent myopic LASIK between July 2002 and May 2003 were analyzed. The two measurements were assessed preoperatively and postoperatively at 1 day and 1, 3, 6, and 12 months. RESULTS: In 237 eyes, using the correction factor 0.93, no significant difference was noted in the preoperative central corneal thickness measured by ultrasonic pachymetry (561.89 +/- 28.66 microm) and Orbscan pachymetry (562.28 +/- 28.18 microm) (P =.713). Postoperatively, the difference was statistically significant at day 1 and 1, 3, and 6 months (P < .001), but was not significant at 12 months (P = .130). CONCLUSIONS: Orbscan II measurements of central corneal thickness after myopic LASIK are less than those measured by ultrasonic pachymetry; however, this difference decreases with time and may not be significant after 1 year.
机译:目的:使用超声测厚法和Orbscan II测量来比较近视LASIK手术后中央角膜厚度,并评估这些测量值随时间的变化。方法:分析了2002年7月至2003年5月接受近视LASIK手术的患者通过超声测厚法和Orbscan II(Bausch and Lomb,Rochester,NY)获得的中央角膜厚度测量值。分别在术前和术后1天,1、3、6和12个月评估两次测量。结果:在237眼中,使用校正因子0.93,通过超声测厚仪(561.89 +/- 28.66微米)和Orbscan测厚仪(562.28 +/- 28.18微米)测量的术前中央角膜厚度没有显着差异(P =。 713)。术后第1天,第1天,第3天和第6个月的差异有统计学意义(P <.001),但在第12个月时无显着差异(P = .130)。结论:Orbscan II测量的近视LASIK术后中央角膜厚度小于超声测厚法。但是,这种差异会随着时间的推移而减小,并且在1年后可能并不明显。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号