Analytical method for localizing a fluorescent inclusion in a turbid medium

机译:在混浊介质中定位荧光夹杂物的分析方法

获取原文
获取原文并翻译 | 示例
           

摘要

We describe a novel method for localizing a fluorescent inclusion in a homogeneous turbid medium through the use of time-resolved techniques. Based on the calculation of the mean time of the fluorescence curves, the method does not require a priori knowledge of either the fluorescence lifetime or the mean time of the instrument response function since it adopts a differential processing approach. Theoretical expressions were validated and experiments for assessing the accuracy of localization were carried out on liquid optical phantoms with a small fluorescent inclusion. The illumination and detection optical fibers were immersed in the medium to achieve infinite medium geometry as required by the model used. The experimental setup consisted of a time-correlated single-photon counting system. Submillimeter accuracy was achieved for the localization of the inclusion.
机译:我们描述了一种通过使用时间分辨技术在均质混浊介质中定位荧光包裹体的新方法。基于荧光曲线平均时间的计算,由于该方法采用差分处理方法,因此不需要先验荧光寿命或仪器响应函数的平均时间。验证了理论表达,并在带有小荧光包裹体的液体光学体模上进行了评估定位准确性的实验。将照明和检测光纤浸入介质中,以实现所用模型所要求的无限几何形状。实验装置包括一个与时间相关的单光子计数系统。对于夹杂物的定位,达到了亚毫米级的精度。