首页> 外文期刊>Applied optics >Simultaneous determination of refractive index and thickness of moderately thick plane-parallel transparent glass plates using cyclic path optical configuration setup and a lateral shearing interferometer
【24h】

Simultaneous determination of refractive index and thickness of moderately thick plane-parallel transparent glass plates using cyclic path optical configuration setup and a lateral shearing interferometer

机译:使用循环路径光学配置设置和横向剪切干涉仪同时测定中等厚度的平面平行透明玻璃板的折射率和厚度

获取原文
获取原文并翻译 | 示例
           

摘要

We present a simple technique for simultaneous determination of refractive index and thickness of moderately thick plane-parallel transparent glass plates (GPs) using a cyclic path optical configuration (CPOC) setup and a wedge shear plate as lateral shearing interferometer. The CPOC setup is used to simultaneously focus the counterpropagating converging beams at a common point at its hypotenuse arm. The apparent thickness and real thickness of the test GP are determined by observing three retrocollimation positions of the GP surfaces with respect to the common focus point. The RI is obtained by dividing the real thickness with apparent thickness of the GP. Presented in this paper are the results obtained for a test GP with a thickness of 14.983 mm and a RI of 1.515.
机译:我们提出了一种简单的技术,可同时使用循环路径光学配置(CPOC)和楔形剪切板作为横向剪切干涉仪,同时确定中等厚度的平面平行透明玻璃板(GPs)的折射率和厚度。 CPOC设置用于将反向传播的会聚光束同时聚焦在其斜边臂的公共点。通过观察GP表面相对于公共焦点的三个后准直位置来确定测试GP的表观厚度和实际厚度。 RI是通过将实际厚度除以GP的表观厚度而获得的。本文介绍的是厚度为14.983毫米,RI为1.515的测试GP的结果。