首页> 外文期刊>Electroanalysis >A Calibration-Base Method for the Evaluation of the Detection Limit of an Electrochemical Biosensor
【24h】

A Calibration-Base Method for the Evaluation of the Detection Limit of an Electrochemical Biosensor

机译:电化学生物传感器检出限评估的基于校准的方法

获取原文
获取原文并翻译 | 示例
           

摘要

The metliod based on the confidence limits of a regression straight line is applied to the linear portion of the calibration curve of a biosensor in order to determine the limit of detection. A practical application gives a detection limit of an ethanol biosensor greater than that calculated by using the method of the signal-to-noise-ratio, SIN = 3, but the value is more reliable because it accounts for the variability in the response at analyte additions. A simple approximation of the method is also presented.
机译:将基于回归直线的置信度极限的方法应用于生物传感器校准曲线的线性部分,以确定检测极限。实际应用中,乙醇生物传感器的检出限大于使用信噪比方法计算的检出限SIN = 3,但该值更可靠,因为它考虑了分析物响应的可变性补充。还介绍了该方法的简单近似。