Reproducibility for isometric and isokinetic maximum knee flexion and extension measurements using the IsoMed 2000-dynamometer

机译:使用IsoMed 2000测力计进行等距和等速最大膝关节屈伸测试的可重复性

获取原文
获取原文并翻译 | 示例
           

摘要

The present study was conducted to evaluate the relative and absolute reproducibility of the IsoMed 2000-dynamometer in measuring Peak Torque (PT) during maximum isometric and isokinetic (60 and 120deg/s) knee flexion and extension. Thirty four physically active male subjects (mean age: 23.2 years) were measured in three sessions (T1-T3), 48-72 h apart. Repeated measures analysis of variance (ANOVA) with Bonferroni post hoc adjustments revealed significant systematic errors for isokinetic flexion measurements with mean values on the second and third visit being around 3-5% higher than the corresponding values obtained in the first visit. Intraclass correlation coefficients (ICC 2,1) of 0.90-0.98 and 0.94-0.98 combined with values of standard error of measurement (SEM) of 5.5-9.1 and 4.0-9.0 Nm were found for T1-T2 and T2-T3, respectively. Based on the systematic errors found for flexor measurements and a clear improvement in reproducibility parameters from T1-T2 to T2-T3 for nearly all measurements, we recommend the use of a familiarisation session prior to actual testing.
机译:进行本研究以评估IsoMed 2000测力计在最大等距和等速(60和120deg / s)膝盖屈伸过程中的峰值扭矩(PT)的相对和绝对再现性。在间隔48-72小时的三个时段(T1-T3)中测量了34位身体活跃的男性受试者(平均年龄:23​​.2岁)。进行Bonferroni事后调整的方差重复测量(ANOVA)分析显示,等速屈曲测量的系统误差显着,第二次和第三次就诊的平均值比第一次就诊得到的平均值高约3-5%。发现T1-T2和T2-T3的类内相关系数(ICC 2,1)分别为0.90-0.98和0.94-0.98,再加上标准测量误差(SEM)值为5.5-9.1和4.0-9.0 Nm。基于针对屈肌测量发现的系统误差以及几乎针对所有测量的从T1-T2到T2-T3的可重复性参数的明显改善,我们建议在实际测试之前使用熟悉的课程。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号