In a patient with a defibrillator system incorporating a lead with integrated bipolar sensing, physical contact between a guidewire (inserted for venous access) and the right ventricular defibrillation coil created sensing artefacts that triggered an inappropriate discharge. The presence of the guidewire in the heart during the discharge resulted in an electrical short with irreversible damage to the generator and rendered it incapable of delivering further high voltage therapy.
机译:对于具有集成了双极感应功能的导线的除颤器系统的患者,导丝(插入静脉通道)和右心室除颤线圈之间的物理接触会产生感应伪像,从而触发不适当的放电。放电期间心脏内存在导丝会导致电气短路,对发生器产生不可逆转的损害,并使其无法进行进一步的高压治疗。