Induction ovens have been reported to exert electromagnetic interference on implanted cardiac pacemakers. In an attempt to quantitatively investigate the electromagnetic interference caused by an induction oven on implantable unipolar cardiac pacemakers, we measured the distribution profile of the magnetic field intensity, both with and without a pan on the induction oven. We also performed the inhibition test and asynchronous test using four kinds of pacemakers housed in the standardized Irnich human body model, and measured the maximal distance from the induction oven up to which the interference occurred. In the pan-detection mode of the oven in the absence of a pan, the distribution profile of the magnetic field intensity peaked at the center of the cooking plate, and during induction heating of a pan placed on the induction oven, it was the largest at the circular top-edge of the pan. Pacemaker pulses were inhibited by the induction oven, or generated by the reversion mechanism. The maximal interference distance from the oven was 34 cm for one of the pacemakers. Thus, the safe distance from an induction oven of a patient with an implanted cardiac pacemaker is considered to be 50 cm or more. In conclusion, in the pan-detection mode of the oven in the absence of a pan, the distribution profile of the magnetic field intensity peaked at the center of the cooking plate, and during the induction heating of a pan placed on the oven, it peaked at the circular edge of the pan. The induction oven asynchronized or generated pulses in implantable unipolar cardiac pacemakers up to a maximal distance of 34 cm from the induction oven.
机译:据报道,电磁炉会对植入的心脏起搏器产生电磁干扰。为了定量研究感应炉对植入式单极心脏起搏器造成的电磁干扰,我们测量了感应炉上有无锅时磁场强度的分布曲线。我们还使用标准化Irnich人体模型中的四种起搏器进行了抑制测试和异步测试,并测量了从感应烤箱到发生干扰的最大距离。在没有锅的烤箱的锅检测模式下,磁场强度的分布曲线在烹饪板的中心达到峰值,并且在感应烤箱上放置的锅进行感应加热时,磁场强度分布最大。在平底锅的圆形顶部边缘。起搏器脉冲被感应烘箱抑制,或由回复机制产生。其中一个起搏器与烤箱的最大干扰距离为34厘米。因此,与具有心脏起搏器的患者的感应炉的安全距离被认为是50cm以上。总之,在没有锅的烤箱的锅检测模式下,磁场强度的分布曲线在烹饪板的中心达到峰值,并且在放置在烤箱上的锅的感应加热过程中,在锅的圆形边缘达到峰值。感应炉在植入式单极心脏起搏器中使脉冲同步或产生脉冲,最远距离感应炉34厘米。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号