首页> 外文会议>Meeting of the Electrochemical Society >Selective Dissolution of Ni from Nitinol for Increasing the Biocompatibility
【24h】

Selective Dissolution of Ni from Nitinol for Increasing the Biocompatibility

机译:从镍钛诺选择性溶解Ni,以增加生物相容性

获取原文

摘要

Different electrochemical surface treatments were applied to the Ni-Ti shape memory alloy and to pure Ti and Ni for comparison. Electrolytes containing the halogen ions F{sup}- and Cl{sup}- produced localized corrosion on Ni-Ti. However, by Rutherford backscattering spectroscopy and scanning electron microscopy a loss of nickel and morphological modifications of the surface were verified for Ni-Ti polarized in 1M H{sub}3PO{sub}4 or 1M Na{sub}2SO{sub}4. Corrosion resistance improvement in Hank's solution was observed however, for the Ni-Ti alloy polarized at 3V in 1M H{sub}3PO{sub}4.
机译:将不同的电化学表面处理应用于Ni-Ti形状记忆合金,并进行纯Ti和Ni以进行比较。含有卤素离子f {sup} - 和cl {sup} - 在ni-ti上产生的局部腐蚀的电解质。然而,通过Rutherford反向散射光谱和扫描电子显微镜,验证了表面偏振的镍和形态修饰的损失,以1M H {Sub} 3Po {Sub} 4或1M Na {Sub} 4偏振。然而,观察到汉克溶液的耐腐蚀性改善,对于在1M H {} 3PO {Sub} 4中以3V偏振的Ni-Ti合金。