首页> 外文会议>V Latin American Congress on Biomedical Engineering >Es Ventajoso Adaptar el Controlador PI al Tipo de Electrodo en Ablacion Cardiaca Percutanea por RF?: Estudio Basado en Modelos Teoricos
【24h】

Es Ventajoso Adaptar el Controlador PI al Tipo de Electrodo en Ablacion Cardiaca Percutanea por RF?: Estudio Basado en Modelos Teoricos

机译:通过RF在经皮心脏烧蚀中适应电极类型的PI控制器适应电极类型是有利的?:基于MoriC模型

获取原文

摘要

The commercially available generators for radio-frequency (RF) cardiac catheter ablation employ a propor-tional-integral (PI) controller. Its performance is based on the value of two parameters, K_p and K_i, which are fixed, regardless the electrode type, i.e. different diameter, length, and tempera-ture sensor type. Our aim was to assess by means of theoretical models whether values of K_p and K_i specifically chosen for each electrode type could affect the performance of the tempera-ture-controlled RF cardiac ablation. We used theoretical models solved by the Finite Element Method. The results suggest that the specific choice of K_p and K_i parameters adapted to different types of electrode does not provide a clinically impor-tant enhancement.
机译:用于射频(RF)心脏导管消融的市售发电机采用助致 - 整体(PI)控制器。其性能基于两个参数,K_P和K_I的值,无论电极类型,即不同的直径,长度和温度传感器类型。我们的目的是通过理论模型进行评估,无论是针对每个电极类型专门选择的K_P和K_I的值是否会影响温度控制的RF心脏消融的性能。我们使用了有限元方法解决的理论模型。结果表明,适用于不同类型电极的K_P和K_I参数的具体选择不能提供临床上的增强。