首页> 外文期刊>BJOG: an international journal of obstetrics and gynaecology >Limitations of ST analysis in clinical practice: three cases of intrapartum metabolic acidosis
【24h】

Limitations of ST analysis in clinical practice: three cases of intrapartum metabolic acidosis

机译:ST分析在临床实践中的局限性:三例分娩期代谢性酸中毒

获取原文
获取原文并翻译 | 示例
           

摘要

I wish to comment on the recent article published in the BJOG on the limitations of ST analysis in clinical practice. All three cases described in the article had a poor neonatal outcome despite the labours being monitored with ST waveform analyser (STAN). The authors suggest that this fact highlights the limitations of intrapartum monitoring with STAN. However, closer scrutiny of the cases clearly show that the reason for the poor outcomes is not the STAN technology itself but the fact that STAN guidelines were repeatedly breached in all the three cases.
机译:我想评论一下BJOG最近发表的关于ST分析在临床实践中的局限性的文章。尽管使用ST波形分析仪(STAN)监控了分娩,但文章中描述的所有三例新生儿的结局均较差。作者认为,这一事实凸显了用STAN进行产时监测的局限性。但是,对这些案件的仔细检查清楚地表明,结果不佳的原因不是STAN技术本身,而是在所有这三个案件中屡屡违反STAN准则的事实。